Health Canada — Medical Devices · Canada · 2018-01-08 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Reported risk: Medtronic is issuing this notification following reports that the Signia Power Handle may shut down unexpectedly during use. The Signia Power Handle Battery Management System may incorrectly report the battery relative state of charge, which can result in a display showing more battery life than is available. In some cases, the power handle may have insufficient energy to complete the intended operation. Use of the affected product may result in the need to use the manual retraction tool to remove the device from tissue.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.