LIFEPAK 20e DEFIBRILLATOR/MONITOR (2017-12-07) — recall 57962
Health Canada — Medical Devices · Canada · 2018-01-08 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Physio-control is aware that some devices have had power-related failures as customers prepared their device for initial deployment or during use within the first year of distribution. The symptoms of these failures may include unexpected power on and power off, device lock-up, or a failure to power on or off, any of which has the potential to result in a failure to deliver therapy to the patient and serious injury or death. These failures are the result of manufacturing process residue located beneath a component mounted on the Power Printed Circuit Board Assembly (PCBA). There have been no adverse events reported as a result of this issue.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.