Immulite System- IGF-1 (Insulin like Growth Hormone) — recall 57732
Health Canada — Medical Devices · Canada · 2018-01-08 · Type III
As described by Health Canada — Medical Devices: Immulite System- IGF-1 (Insulin like Growth Hormone) - 1000 and 2000 Assay (2017-11-30)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: In response to customer feedback, Siemens evaluated alternate statistical methods to determine if an alternate method would be a better fit for the reference range data published in the Information for Use (IFU). Our investigation concluded that the LMS method provided a better fit to the data and produced ranges more consistent with the expected physiological change in IGF-I levels compared to use of the weighted average results reported in the IFU.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.