Product Wiki · Recalls · Health Canada — Medical Devices

Disposable Drill Kit (2017-12-21) — recall 57616

Health Canada — Medical Devices · Canada · 2018-01-12 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
57616
Date of alert
2018-01-12
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: This recall has been initiated due to a drill bit mix-up. DDK2-2.4-30X and DDK2-2.8-30X Disposable Drill Kits were found to be packaged at the same time causing devices to be labeled incorrectly.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/disposable-drill-kit-2017-12-2

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.