Sorin Centrifugal Pump CP5 System and Control Panel (2017-10-27) — recall 57574

Health Canada — Medical Devices · Canada · 2017-11-10 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
57574
Date of alert
2017-11-10
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: During internal testing, LivaNova identified a possibility that automatic line closure by The Electrical Remote-Controlled Tubing Clamp (ERC) after detection of an air bubble could take longer than is intended by design in the CP5 system. This could lead to the remote possibility of an air bubble reaching a patient before the clamp closes if the blood flow rate is high and the distance between the bubble sensor and the patient is set at the minimum required by the IFU (1 meter).
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/sorin-centrifugal-pump-cp5-sys

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.