PLUM 360 INFUSION SYSTEM WITH MEDNET (2017-10-19) — recall 57573
Health Canada — Medical Devices · Canada · 2017-11-10 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Issue 1: The user is not able to turn off the pump when a pump in a distal occlusion paused state is exposed to a second "power off then on" malfunction. Issue 2: When the user accesses the preventive maintenance screen, in service/biomed mode, on Plum 360 v15.10.00.010 devices, with a total delivery time greater than 1,500 hours, the user will not be able to interact with the device and the device must be power cycled.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.