Health Canada — Medical Devices · Canada · 2017-11-10 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: On October 6, Southmedic received a customer complaint in which the CO2 diffuser of the EtCO2 mask became disconnected from the mask and fell into the patient's mouth. Further inspection of the identified lot number and investigation into root cause identified the potential for loose or missing diffusers in other lots.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.