Various Siemens Dimension Systems (2017-10-20) — recall 57537
Health Canada — Medical Devices · Canada · 2017-10-30 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: The biotin concentration in the non-interfering table for dimension FT4L and TSHL and Dimension Vista TSH is incorrect and the biotin concentration is being revised. In addition, Dimension Vista CTNI and MMB and Dimension SIRO do not indicate biotin interference in the non-interfering table of the IFU. The IFU will be updated with the established claim for biotin.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.