Product Wiki · Recalls · Health Canada — Medical Devices

BC-THROMBIN REAGENT (2017-10-12) — recall 57531

Health Canada — Medical Devices · Canada · 2017-10-27 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
57531
Date of alert
2017-10-27
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: The BC Thrombin Reagent Kit lots 46751 (contains Thrombin Reagent lot 517468) and Kit lot 47184 (contains Thrombin Reagent lot 517469) produce unexpected prolonged thrombin time results for expected normal samples, and may recover above the upper limit of normal which is 21 sec as stated within the IFU (instruction for use) and the BCS/ BCS XP System application sheets.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/bc-thrombin-reagent-2017-10-12

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.