IVEA MODEL 500A (2017-10-11) — recall 57528

Health Canada — Medical Devices · Canada · 2017-10-27 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
57528
Date of alert
2017-10-27
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Rotational forces on the 16082 dowel pin during deploying or folding can cause the pin to slowly "screw" its way out of position. If the pin migrates enough to allow the brace to detach from the leg pivot fitting while a patient is ambulating with weight on the handles, the product could slump to one side. This may cause the patient to stumble.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/ivea-model-500a-2017-10-11

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.