Product Wiki · Recalls · Health Canada — Medical Devices

Cobas Liat Analyzer — recall 57485

Health Canada — Medical Devices · Canada · 2017-09-05 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
57485
Date of alert
2017-09-05
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: The Cobas Liat System allows the use of several different barcode types for sample identification. If any of these barcode types are used together with Codabar or Code 39 barcodes and match an existing Codabar or Code 39 exactly, it is theoretically possible that a sample mismatch between the two samples could occur when using software version 3.0.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/cobas-liat-analyzer

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.