Renew Reusable Scissor, Grasper And Dissector Tips — recall 57417
Health Canada — Medical Devices · Canada · 2017-08-15 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Microline Surgical, inc., has initiated this global voluntary recall which applies to all renew reusable tips (scissors, graspers, and dissectors). This recall is due to a malfunction related to the heat shrink insulation tube that is located immediately proximal to the metal tip. Microline Surgical, has received customer reports that the heat-shrink insulation tube may split during autoclave (sterilization/re-sterilization), and could fall-off during surgery, and if cautery is used, there is potential for burn.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.