Product Wiki · Recalls · Health Canada — Medical Devices

Au Lyophilized Chemistry Calibrator (2017-08-28) — recall 57293

Health Canada — Medical Devices · Canada · 2017-09-26 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
57293
Date of alert
2017-09-26
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Unpaired lots of au lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. Paired calibrator lots are identified by the last two digits of the lot number. If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/au-lyophilized-chemistry-calib

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.