Triathlon Total Knee System - Ts Femoral Component — recall 57255
Health Canada — Medical Devices · Canada · 2017-07-18 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: It was discovered during manufacturing, that size 3 and 4 Triathlon Cemented Revision Femoral components had excess metal trapped in the size and lot code marking imprints. Analysis showed that the material trapped in the imprint was 304 stainless steel blast media, which is used to grit blast the implants during the manufacturing process.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.