Health Canada — Medical Devices · Canada · 2017-03-31 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Beckman Coulter has confirmed through internal investigation that the access reaction vessel (RV) lots listed fail to meet a specification for outside diameter on the open ended side of the RV. - The RV lots listed may cause aliquot vessel (AV) shuttle motion errors when they are loaded onto the Unicel DXC 600i Synchron Access Clinical System through the closed tube aliquotter (CTA). There is no impact to patient results. - Internal testing did not demonstrate shuttle motion errors on the access 2 standalone system.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.