XN-10 and X-11 Modules (2017-04-04) — recall 56838
Health Canada — Medical Devices · Canada · 2017-05-02 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: The protective plastic sheet that is intended to cover and protect electronic components inside the XN-series analyzers was not positioned correctly during manufacturing. Reagent from a leaking valve created a build-up of salt on an electronic component because of the misalignment of the protective sheet. The salt build-up created a salt bridge that allowed current to continue to flow, create heat, and eventually ignite adjacent resin parts inside the analyzer.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.