Cook Bush Ureteral Illuminating Catheter Set — recall 56751
Health Canada — Medical Devices · Canada · 2017-04-04 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: The proximal black component that connects to the distal transparent component of the catheter may overheat and melt. There have been no reports of overheating along the transparent component of the catheter, which is the part of the device that comes into contact with the patient's urethra, bladder, and ureter. Potential adverse events that may occur if these devices overheat and come into direct contact with skin include burns to the skin.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.