Carestation 620 and 650 (2017-05-02) — recall 56633
Health Canada — Medical Devices · Canada · 2017-06-12 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: GE Healthcare has recently become aware of a potential safety issue with the CareStation 620, 650 and 650C. An incomplete seal can exist between the disposable absorber and the breathing circuit lower assembly of the CareStation 600 series systems. This incomplete seal can allow rebreathing of patient gases that have bypassed the carbon dioxide (CO2) absorbent material and could result in unintended elevated inspired levels of co2 (FICO2), which could lead to hypercarbia.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.