Reported risk: Philips discovered that some welding joints on the vertical carriage brackets of the tube stand may crack. As a result, the tube arm assembly may exhibit uncharacteristic sideways movement, with the tube becoming off center. If all of the welding joints cracked, the tube arm assembly could fall down.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.