Product Wiki · Recalls · Health Canada — Medical Devices

StrataMR Valve (2017-03-06) — recall 56560

Health Canada — Medical Devices · Canada · 2017-03-16 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
56560
Date of alert
2017-03-16
Country / jurisdiction
Canada
Risk level
Type I
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Medtronic is initiating this voluntary recall because the StrataMR Valve may experience a condition that may, in limited circumstances, affect the flow resistance in the valve and result in underdrainage of cerebrospinal fluid (CSF). The recall is based on an issue impact assessment, where an increase in the complaint rate related to incorrect flow was noted for the StrataMR product family which indicated that current residual risk for StrataMR products is undesirable and requires additional mitigations to reduce risk. This condition may result in the following adverse health consequences: headaches, nausea, vomiting and lethargy. If left untreated underdrainage has the potential to lead to coma and death.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/stratamr-valve-2017-03-06

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.