Central Water Plant Reverse Osmosis Units (CWP 100 and 60)… — recall 56476
Health Canada — Medical Devices · Canada · 2016-12-05 · Type II
As described by Health Canada — Medical Devices: Central Water Plant Reverse Osmosis Units (CWP 100 and 60) (2016-11-23)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Baxter Corporation is issuing a device correction for specific models of the central water plant (CWP) reverse osmosis units in order to correct the insufficient crimping of the cables within the main pump. The issue was identified as a result of customer complaints received for the presence of smoke in the water treatment room due to the overheating of the pump in the CWP. As part of Baxter's investigation, based on pictures provided from the event, the most likely cause of the thermal event was due to the crimp of the cables for the main pump being insufficient. To minimize any potential side effects of thermal events, please ensure all panels of the CWP are in place.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.