Product Wiki · Recalls · Health Canada — Medical Devices

Arkon Anesthesia System — recall 56343

Health Canada — Medical Devices · Canada · 2017-03-16 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
56343
Date of alert
2017-03-16
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Spacelabs has received reports of the failure of the power supply that provides electricity to the Arkon Subsystems. This failure causes the system to run on its internal battery(s) without charging them or providing the user the "on battery" display messages. Once the batteries reach their minimal charge level, the Arkon powers down and the backup audible buzzer alarm activates.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/arkon-anesthesia-system

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.