Biolox Forte 36mm Alumina Ceramic Femoral Heads — recall 56285
Health Canada — Medical Devices · Canada · 2017-03-10 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: It was found during Microport's Investigation that the Biolox Forte 36mm Alumina Ceramic Head was confirmed to measure approximately 2mm shorter than other Microport 36mm Femoral Heads (Cobalt Chrome and Biolox Delta Ceramic Femoral Heads). The ultimate hazard would be joint laxity and/or dislocation, which could result in revision surgery.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.