Product Wiki · Recalls · Health Canada — Medical Devices

Vitros Chemistry Products Na+ Slides — recall 56094

Health Canada — Medical Devices · Canada · 2017-02-01 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
56094
Date of alert
2017-02-01
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Ortho Clinical Diagnostics (Ortho) identified the potential for biased results to be generated over the 10 day on-analyzer storage limit when using VITROS Na+ cartridges warmed between 1.5 to 8 hours then placed on the analyzer. Ortho has determined that cartridges require a minimum of 8 hours of warm up at room temperature to help ensure the slides have acceptable performance throughout the 10 day on-analyzer storage limit.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/vitros-chemistry-products-na-s

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.