Product Wiki · Recalls · Health Canada — Medical Devices

ACTIVA Neurostimulator Charging Systems (2016-11-01) — recall 56068

Health Canada — Medical Devices · Canada · 2016-11-18 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
56068
Date of alert
2016-11-18
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Medtronic has identified an increased number of complaints from customers involving reports of rechargers that are in an unresponsive error state, where the recharger is non-functional with a blank display screen and is beeping every 5 seconds. When this error state occurs, the recharger is unable to recharge the neurostimulator until the recharger is reset. If the neurostimulator battery is allowed to become fully depleted, this can lead to loss of therapy and return of associated disease-specific symptoms. If the implanted neurostimulator battery is allowed to remain fully depleted, it may overdischarge, resulting in a permanent reduction in battery capacity and the need to recharge more frequently in the future. For model 37751 Implantable Neurostimulator (INS) rechargers that are used by patients to recharge their INS devices, notification will be provided to active implanting and managing physicians of both spinal cord stimulation (SCS) and deep brain stimulation (DBS) therapies. All model 37751 rechargers that are included in model 37651, 37754 and 97754 kits may be affected by this issue. For a subset of patients receiving DBS therapy, in rare instances, a loss of DBS therapy may result in a life threatening injury or death. For example, patients being treated for Parkinson's Disease may experience akinetic crisis, and patients treated for epilepsy may experience status epilepticus. Medtronic has not received any reports of life threatening injury or death associated with this issue.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/activa-neurostimulator-chargin

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.