Brilliance 40/64 Slice CT System, Ingenuity CT System (2016-08-29) — recall 55990

Health Canada — Medical Devices · Canada · 2016-08-29 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
55990
Date of alert
2016-08-29
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Philips has become aware that while performing high resolution head scans, the reconstructed images may exhibit degraded image quality manifested as: non uniformity of the image CT number shift, greater than 5 Hounsfield units reduction in gray/white matter differentiation the issue is most severe in the high resolution mode. Philips provides factory reference pediatric scan protocols which are set to high resolution by default. Adult head scans have a factory reference protocol that defaults to standard resolution which does not experience the issue
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/brilliance-4064-slice-ct-syste

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.