Product Wiki · Recalls · Health Canada — Medical Devices

LIFEPAK 15 MONITOR/DEFIBRILLATOR — recall 55628

Health Canada — Medical Devices · Canada · 2016-09-09 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
55628
Date of alert
2016-09-09
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Physio-Control, Inc. LIFEPAK 15 Monitor/Defibrillators with End-tidal(Et)CO2 capability are distributed with the EtCO2 setting preset to display in (millimeter of mercury) mmHg. Customers may choose to reconfigure the device to display the EtCO2, in alternative modes of (kilopascal) kPa or percent settings. When using EtCO2 in the kpa or percent setting and in a situation where the reading is above 9.9 kPa, the display of the LIFEPAK 15 respiratory rate may partially obscure a portion of the leading digit of the EtCO2 value. Although part of the first digit of the EtCO2 value is obscured by the respiratory rate, the capnography waveform scale automatically adapts to the EtCO2 value. Observing the capnography waveform and the range displayed will help the user confirm the patient's EtCO2 value. Clinical interpretation of a low EtCO2 value when the value is actually high could potentially lead to incorrect treatment. There have been no reports of incorrect patient treatment or harm associated with this issue.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/lifepak-15-monitordefibrillato

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.