Dimension vista System-Chem calibrators — recall 55600
Health Canada — Medical Devices · Canada · 2016-09-20 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Siemens Healthcare Diagnostics has confirmed that DIMENSION VISTA CHEM 1 CAL Lots 5GM081 and 5GM082 may produce depressed results with current lots of calcium reagent. Although the affected CHEM 1 CAL lots are expired, the calcium calibration interval is 90 days, therefore it is possible that calibrations using the calibrator lots are active.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.