Health Canada — Medical Devices · Canada · 2016-07-22 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: The current IOLmaster 700 software version 1.14.84 sometimes shows an unexpected behavior during toric IOL power calculation with the haigis suite: in some cases of low cylinder power with a theoretically ideal IOL below the delivery range of the selected toric IOL model, the device suggests a toric IOL with significant remaining cylinder refraction.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.