REVOLUTION CT - MAIN UNIT (2016-04-21) — recall 55366

Health Canada — Medical Devices · Canada · 2016-05-16 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
55366
Date of alert
2016-05-16
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: GE Healthcare has become aware that an out of tolerance torque tool was used on three joints on revolution CT system during the manufacturing process. If the x-ray tube becomes loose on the CT gantry during operation, this could result in beam tracking errors, potential scan aborts and/or a possible image artifact. There is a secondary safety mechanism to prevent components from being expelled from the gantry if the bolts fail. There have been no injuries reported as a result of this issue.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/revolution-ct-main-unit-2016-0

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.