Product Wiki · Recalls · Health Canada — Medical Devices

ADVIA Centaur XPT Instrument (2016-06-21) — recall 55146

Health Canada — Medical Devices · Canada · 2016-06-30 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
55146
Date of alert
2016-06-30
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Siemens Healthcare Diagnostics is issuing a customer notification for ADVIA Centaur XPT Waste Reservoir Assembly (SMN 10711091). Siemens has determined that certain reservoirs manufactured with walls that are below the nominal specification, potentially cause cracking & leaking of seam bottom and leaking at the interface of the seam and reservoir. If leak occurs foaming may occur in the glass jar in the bulk fluid drawer and cause the instrument to stop with low vacuum error #016001324.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/advia-centaur-xpt-instrument-2

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.