AISYS CS2 ANESTHESIA MACHINE MAIN UNIT (2016-02-18) — recall 54988
Health Canada — Medical Devices · Canada · 2016-03-04 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: GE Healthcare has recently become aware of a potential safety issue involving a momentary, self-correcting anesthetic agent bolus when using 21% oxygen on all Aisys CS2 and upgraded Aisys anesthesia devices. Delivery of this momentary bolus of anesthetic agent is potentially hazardous because it could lead to hypotension in certain vulnerable pediatric patients when 21% oxygen (air only) is used. There have been no injuries reported as a result of this issue.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.