Product Wiki · Recalls · Health Canada — Medical Devices

AISYS CS2 ANESTHESIA MACHINE MAIN UNIT (2016-02-18) — recall 54988

Health Canada — Medical Devices · Canada · 2016-03-04 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
54988
Date of alert
2016-03-04
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: GE Healthcare has recently become aware of a potential safety issue involving a momentary, self-correcting anesthetic agent bolus when using 21% oxygen on all Aisys CS2 and upgraded Aisys anesthesia devices. Delivery of this momentary bolus of anesthetic agent is potentially hazardous because it could lead to hypotension in certain vulnerable pediatric patients when 21% oxygen (air only) is used. There have been no injuries reported as a result of this issue.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/aisys-cs2-anesthesia-machine-m

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.