Product Wiki · Recalls · Health Canada — Medical Devices

ZILVER VASCULAR STENT WITH RAPID EXCHANGE (2016-02-09) — recall 54977

Health Canada — Medical Devices · Canada · 2016-02-26 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
54977
Date of alert
2016-02-26
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Testing demonstrates a process used to manufacture the inner catheter to stylet joint of the delivery system doesn't meet the required specification on a consistent basis and there is a risk the joint may separate during deployment. A failure of this joint would represent a device malfunction and were it to occur during a procedure may result in a partial stent deployment. The clinical effect of a partial deployment may be a need for surgical intervention, or may result in vessel trauma or an embolic event
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/zilver-vascular-stent-rapid-ex

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.