Lightspeed 16 CT Scanner and Discovery PET/CT 600 Systems (2016-02-05) — recall 54901
Health Canada — Medical Devices · Canada · 2016-02-12 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: GEHC has updated the service manuals used to maintain these scanners to fix an incorrect torque value, which is the source of the field action. FMI 25463 identified a torque issue with 5 volt supplies in CT and PET LightSpeed systems and corrected the issue by field action, as well as a technical manual update. This recall is to send a second customer letter to anyone who may still possess and use an obsoleted service manual.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.