ADVIA CENTAUR SYSTEMS (2015-12-30) — recall 54829

Health Canada — Medical Devices · Canada · 2016-01-22 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
54829
Date of alert
2016-01-22
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Not meeting the current High-Dose Hook Effect expectation in the Instructions For Use (IFU). The ADVIA Centaur systems PSA assay range is 0.01-100 ng/mL (μg/L). The ADVIA Centaur Systems PSA IFU states that patient samples with high total PSA levels can cause a paradoxical decrease in the Relative Light Units (RLUs) (high-dose hook effect). In this assay, patient samples with total PSA levels as high as 10,000 ng/mL (μg /L) will assay greater than 100 ng/mL (μg/L).
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/advia-centaur-systems-2015-12-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.