ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEMS AND STRIPS… — recall 54811
Health Canada — Medical Devices · Canada · 2016-02-01 · Type II
As described by Health Canada — Medical Devices: ICHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEMS AND STRIPS (2016-01-11)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: IRIS international has become aware of loose or missing analyte pads being found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the iChem VELOCITY system. If affected strips are used to assay patient samples, the issue has the potential to cause erroneous pH results and false negative results for the following tests: nitrite, glucose, blood, bilirubin, urobilinogen, protein, leukocyte esterase, ketones, and ascorbic acid.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.