Health Canada — Medical Devices · Canada · 2016-02-08 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Based on discrepant results (false negative results) reported from the field, bioMérieux initiated a complaint investigation. A loss of performance due to light exposure after 3 months of shelf life for API ZYM B, which lead to false negative results on some API strips using the ZYM B (API 20 Strep-20600, API Coryne-20900, API Staph-20500, API ZYM-25200, API NH-10400, API Listeria-10300).
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.