enGen Laboratory Automation System (2016-03-18) — recall 54727

Health Canada — Medical Devices · Canada · 2016-04-07 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
54727
Date of alert
2016-04-07
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Ortho Clinical Diagnostics (Ortho) has initiated this recall pertaining to the enGen Laboratory Automation System configured with Thermo Scientific Recapper Module due to the possibility that the electrical cable that supplies power to the cap Vibratory Feeder of the Recapper Module may loosen from its correct position above the Vibration Feeder support plate and migrate to under the support plate. If this occurs, the uncapped sample tubes may hit the cable, and sample fluid (e.g., serum) may splash from the tube and into another tube or onto the under carriage of the support plate.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/engen-laboratory-automation-sy

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.