Health Canada — Medical Devices · Canada · 2016-01-29 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Radiometer has been made aware that the AQURE system, may display the sample type as "blank", if the sample type is not specified by the user. As multiple result values for the same patient and the same parameter can be displayed side by side on the screen, there is a potential risk of misinterpretation of results, the user may confuse the blank/unspecified sample type with the content of a specified sample type field.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.