DIMENSION VISTA SYSTEM - CHEM CALIBRATORS (2016-03-01) — recall 54607
Health Canada — Medical Devices · Canada · 2016-03-11 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Siemens Healthcare Diagnostics has received complaints of "calibrator insert missing" alerts on Dimension Vista Chemistry 3 Calibrator lot 5md004. Internal investigation has confirmed that the vial barcode label reads KC130 instead of the correct catalog number KC130A. When this mismatch occurs between the vial barcode label and the Instructions For Use (IFU) label, the "calibrator insert missing" error is triggered and displayed on the Dimension Vista instrument screen. There is no risk of erroneous results when this issue occurs as no results are generated.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.