Product Wiki · Recalls · Health Canada — Medical Devices

COULTER DILUENTS (2015-09-04) — recall 54509

Health Canada — Medical Devices · Canada · 2015-10-14 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
54509
Date of alert
2015-10-14
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Beckman Coulter has received an increased number of complaints related to the hemoglobin (Hgb) parameter on the Coulter LH 500 and Coulter HmX Hematology Systems. The investigation has revealed that the issues may be related to variability in a component of the Coulter ISOTON III and Coulter LH Series diluent reagents which in turn may lead to a film coating the inside of the white blood cell (WBC) bath where Hgb is measured. The use of the affected diluent lots may result in symptoms including: Failures in Hgb background at startup or during a background cycle Increases in incomplete computations (...) for Hgb
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/coulter-diluents-2015-09-04

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.