Product Wiki · Recalls · Health Canada — Medical Devices

PHEON MODULAR KNEE JOINTS (2015-07-24) — recall 53941

Health Canada — Medical Devices · Canada · 2015-08-11 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
53941
Date of alert
2015-08-11
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Issue identified where locking function may fail if it is activated if the locking pin is not completely pulled out by patient; edge loading in locking surface causes a deformation of locking pin. The pin may get stuck in the guide of the upper section hindering the lock from engaging automatically in the extended position. The stability of the joint is then only by alignment and polycentric joint design and may result in buckling of the loaded joint.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/pheon-modular-knee-joints-2015

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.