Health Canada — Medical Devices · Canada · 2015-08-10 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Roche received a customer complaint regarding elevated imprecision and too high recovery of STFR control set level 1 and DUP alarms during calibration with the reagent TQ Soluble Transferrin Receptor (STFR). The issue of elevated imprecision was confirmed by internal investigations on lot #606449 on Cobas C501 showed elevated imprecision of up to 8.9%. The other current reagent lots are not affected by this issue
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.