CDI BLOOD PARAMETER MONITORING SYSTEM 500 — recall 53760
Health Canada — Medical Devices · Canada · 2015-09-24
As described by Health Canada — Medical Devices: CDI BLOOD PARAMETER MONITORING SYSTEM 500 - MONITOR WITH BLOOD PARAMETER MODULES AND ONE HCT/SAT PROBE (2015-08-25)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Terumo Cardiovascular Systems (Terumo CVS) has determined that changes made when upgrading CDI system 500 software from version 1.65 to version 1.69 were not expected or fully understood by users, and that the following user instructions need to be clarified: In-vivo calibration: measured values prior to initial in-vivo calibration may not be used for patient management. at the beginning of a case, you must complete calibration of all measurable blood parameters by comparing them to a laboratory measurement done on a blood sample. The values are dimmed on the CDI system 500 screen to indicate that the values are not accurate until an initial in-vivo calibration is performed.Operating ranges: if a lab measured value is outside the operating range for a parameter, you may not be able to adjust the CDI system 500 measured values to the lab measured value. Accuracy has not been established for displayed values outside the operating range. Therefore, displayed values outside the operating range should not be used for patient management. While this condition exists, continue patient management with another source (i.e., laboratory or point-of-care blood gas analyzer). See CDI system 500 operating ranges: reference table for system accuracy limits.Temperature: the temperature measured by the shunt sensor is local to the sensor and does not reflect the actual patient arterial temperature. Do not use this measurement for patient management.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.