Health Canada — Medical Devices · Canada · 2015-08-18 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: As a result of in-house product testing, it was discovered that when performing a 2D stress echo study, specific keystrokes result in a potential loss of data, where some selected clips may not be saved as part of the study. The issue is present only for users with software versions VA16A and VA30A. It is solved in customers with software version 3.5 (VA35) or 4.0(VA40).
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.