Product Wiki · Recalls · Health Canada — Medical Devices

Dimension Vista System (2015-05-20) — recall 53576

Health Canada — Medical Devices · Canada · 2015-06-15 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
53576
Date of alert
2015-06-15
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Issue # 1: Manufacturer received complaints for samples not processing without notification on the Dimension Vista 500 or Dimension Vista 1500 using software versions 3.6.1 in most cases, user action of pause / cancel pause is effective in restoring lane operation and the samples will resume processing. However in a few cases it was necessary to shutdown the instrument software and restart to recover. In all cases the affected samples are resubmitted and run successfully. Issue # 2: Manufacturer received complaints for unexpected less than lower assay range (< l) or results that were flagged with assay errors on Dimension Vista 1500 systems using software versions 3.6.1. Result data and the physical evidence from flexes suggested that a reagent server lost synchronization and the software was unaware of the problem.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/dimension-vista-system-2015-05

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.