SPECTRUM INFUSION WITH MDL VERSION 8.0 (2015-03-06) — recall 53550
Health Canada — Medical Devices · Canada · 2015-04-13 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Baxter has identified a discrepancy between the loading/bolus default dose settings in the master drug library ( MDL ) version 8.0 and the values displayed on the pump during loading/bolus dose programming. When a drug's loading dose or bolus dose is configured in the MDL with a default starting time value in seconds, the loading dose or bolus dose setup screen on the pump will display the time value rounded to the nearest integer in minutes. When this occurs, the pump will administer drugs as configured. This discrepancy is only with the default time on the setup screen, not with the amount or rate of drugs administered, as configured in the drug library.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.