VITROS 5,1 FS CHEMISTRY SYSTEM - VARIOUS COMPONENTS (2015-03-11) — recall 53526

Health Canada — Medical Devices · Canada · 2015-04-14 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
53526
Date of alert
2015-04-14
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Ortho Clinical Diagnostics (OCD) has identified an anomaly with Vitros System software version 3.1 and below. OCD determined that the software may not properly identify an expired calibration. Calibration status is monitored by the system software which is based upon the calibration interval and the date of the last calibration. OCD's investigation confirmed that if the anomaly occurs, a calibration may expire, but the system may not alert the operator and not flag results with a CE sample result code (i.e. calibration expired). The system may unexpectedly generate results even though it is configured to not report results when a calibration has expired.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/vitros-51-fs-chemistry-system-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.