Product Wiki · Recalls · Health Canada — Medical Devices

FEM-FLEX II CANNULA (2015-03-25) — recall 53519

Health Canada — Medical Devices · Canada · 2015-04-27 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
53519
Date of alert
2015-04-27
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Through post market surveillance, Edwards Lifesciences has identified a potential health risk to patients regarding the use of Fem-Flex II femoral arterial cannula, sizes 8, 10, and 12 French only. Edwards has received one customer complaint regarding a released wire, located at the tip area of the cannula, which was identified prior to use. Although the condition does not affect the functionality of the cannula, there is potential patient safety risk if a protruding wire is not detected prior to use.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/fem-flex-ii-cannula-2015-03-25

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.