Reported risk: Sysmex America, Inc. received complaints from Sysmex installation specialists and customers that level 2 of the five levels of white blood cell (WBC) linearity of range check XII (RCXLL) was recovering above the expected range on the POCH-1 OOI, XP-300 and KX-21 n, and KX-21 analyzers (three-part differential analyzers). Level 2 WBC recovery is expected to be approximately 0.5 +/- 0.5 x10 3 /uL and recovery was above 1.0 x10 3 /uL in some instances. When the recovery is above 1.0, the plotted line is not straight suggesting the analyzer is not linear through the entire wbc reportable range of 1.0 to 99.9 x10 3 /uL for the three part differential analyzers. Exclusion of the level 2 produces a linear graph.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.